Laserfiche WebLink
January 8, 2014 Lab ID : STK1352210 <br /> Musco Family Olive Environmental Customer : 3-11544 <br /> Quality Control-Organic <br /> Constituent Method Date/ID Type Units Conc. QC Data DQO Mute <br /> Organic <br /> OC 5310C 01/03/14:200069AMM Blank mg/L ND <0.3 <br /> MS mg/L 40.00 61.3% 75-114 435 <br /> (SP 1313560-006) MSD mg/L 40.00 71.2% 75-114 435 <br /> MSRPD mg/L 40.00 13.5% 1523.0 <br /> 5310C 01/03/14:200108AMM CCV ppm15.00 108% 67-122 <br /> CCV m 15.00 106% 67-122 <br /> Definition <br /> CCV :Continuing Calibration Verification-Analyzed to verify the instrument calibration is within criteria. <br /> Blank :Method Blank-Prepared to verify that the preparation process is not contributing contamination to the samples. <br /> MS Matrix Spikes-A random sample is spiked with a known amount of analyte.The recoveries are an indication of how that sample <br /> matrix affects analyte recovery. <br /> MSD :Matrix Spike Duplicate of MS/MSD pair-A random sample duplicate is spiked with a known amount of analyted.The recoveries <br /> are an indication of how that sample matrix affects analyte recovery. <br /> MSRPD :MS/MSD Relative Percent Difference(RPD)-The MS relative percent difference is an indication of precision for the preparation <br /> and analysis. <br /> ND :Non-detect-Result was below the DQO listed for the analyte. <br /> DQO :Data Quality Objective-This is the criteria against which the quality control data is compared. <br /> Explanation <br /> 435 Sample matrix may be affecting this analyte.Data was accepted based on the LCS or CCV recover . <br /> Pace I I of I I <br />