Laserfiche WebLink
CALIBRATION FORM <br />DATE: February 15, 2016 <br />CUSTOMER: St Joseph's Medical Center <br />ADDRESS: 1909 Cemetery Lane CITY: Stockton <br />STATE: CALIFORNIA 95204 <br />CONTACT: John Hawkins TEL: (209)461-5260 <br />VESSEL NUMBER: 341 TEMPERATURE ACHIEVED: 275 DEGREE F. <br />PRESSURE ACHIEVED: 32 PSIG. THERMOMETER READING: 275 DEGREE F. <br />VALIDATION PERFORMED BY: Shawn Fallan <br />THIS CALIBRATION WAS PERFORMED BY A FACTORY REPRESENTATIVE TO <br />FACTORY SPECIFICATIONS. <br />COMMENTS: All gauizes were checked and are within factor settings. ettings. The autoclave is in <br />good condition with no visible problems. I adjusted the temperature gauge on the control panel to <br />read the correct temperature Autoclave has a lot of pitting around the sealing surface of the door. <br />METHOD OF CALIBRATION: THERMALMETER, FLUKE MODEL 51/52I1, RANGE 100 <br />TO 300 DEGREE'S FAHRENHEIT. THERMALMETER WAS PLACED <br />AT THE FRONT OF THE VESSEL AND SUSPENDED AT 5" ABOVE THE FLOOR. <br />THE CHART RECORDER SET POINT WAS CALIBRATED TO THE TEMPERATURE <br />ACHIEVED BY THE THERMALMETER. <br />THE STERILIZER SHOULD BE RE -CALIBRATED ON OR BEFORE February 15, 2017 <br />1 <br />