! 0 <br />lei <br />CALIBRATION FORM <br />DATE: February 6, 2019 <br />CUSTOMER: St Joseph's Medical Center <br />ADDRESS: 1909 Cemetery Lane CITY: Stockton <br />STATE: CALIFORNIA 95204 <br />CONTACT: Yvonne Serbo TEL: (209)467-6308 <br />VESSEL NUMBER: 341 TEMPERATURE ACHIEVED: 275 DEGREE F. <br />PRESSURE ACHIEVED: 38 PSIG. THERMOMETER READING: 275.9 DEGREE F. <br />VALIDATION PERFORMED BY: Shawn Fallan <br />THIS CALIBRATION WAS PERFORMED BY A FACTORY REPRESENTATIVE TO <br />FACTORY SPECIFICATIONS. <br />COMMENTS: The autoclave is operating properly. Both drain screen were cleaned, door was <br />check for proper alignment, door gasket cleaned, door sealing surface deburred and cleaned, <br />steam traps cleaned out, y -strainers cleaned, inside of the autoclave was also cleaned. Autoclave <br />ran a complete cycle to verify time and temperature. <br />METHOD OF CALIBRATION: THERMALMETER, FLUKE MODEL 51/52II, RANGE 100 <br />TO 300 DEGREE'S FAHRENHEIT. THERMALMETER WAS PLACED <br />AT THE FRONT OF THE VESSEL AND SUSPENDED AT 5" ABOVE THE FLOOR. <br />THE CHART RECORDER SET POINT WAS CALIBRATED TO THE TEMPERATURE <br />ACHIEVED BY THE THERMALMETER. <br />THE STERILIZER SHOULD BE RE -CALIBRATED ON OR BEFORE February 6, 2020 <br />1 <br />