Laserfiche WebLink
Sterilization certified in line with <br />MPI <br />EN ISO 13486:20161 EN ISO 11135:2014 <br />Confirmation of Sterilization <br />Method: VA S_1 <br />Sterilization date: 202011-17 <br />5tedliza0on lot: 18595 <br />Customer. MT.DERM GmbH <br />Your requisition no.: 11112020'^^�r <br />Our delivery note: 253088 from 2020-11-20 <br />Item: 9900031 <br />Sterilisation Pal. MP01 1d NK <br />Pr.70 <br />Quantity: 7 Pal. <br />Nominal values Actual values <br />Incoming time: <br />min. 12 h 96 h <br />Preconditioning min. 16 h 22,75 h <br />Dwell: <br />Conditions: 3&45°C 50-80%rel.htamidily 35.45°C 5080%rel.humidity <br />Sterilization <br />Cycle: C ° °C <br />Temperature: 40 t3 124 mbar abs <br />Prev: 120 mbar abs <br />Humidity: 50-80% <br />rel.humidity 69,9% rel.humidity <br />3h <br />Dwell time: 3h 16-32 kg 26,5 kg <br />EO mass: <br />Airwashes: min. 3 3 <br />Aeration min 1 Days 2,5 Days <br />Dwell: <br />l:�rlainrt 35-45°C 3645°C <br />Remarks: The evaluation fro <br />m the BI's is in process. <br />The physical parameters were in accordance to the specifications. <br />No growth was observed for the mlcrobiologicai inprocess conVol with biolog cal <br />indicators (Bacillus atrophaeus, 3M Attest). <br />The sterilization was conducted in line with the customer validation parameter. <br />go= <br />Signature <br />