Laserfiche WebLink
<br /> EO Gas Sterilization Report <br />Description Lot #Quantity (cartons)Sterilization Date <br />Microblades 3001 360 2024-12 <br /> Process Parameters <br /> Item Unit of Measure Setting Tolerance Range <br />Precondition temperature ℃50 45~55 45~55 <br />Precondition humidity %RHDFDF 40 40~70 40~65 <br />Precondition hold MINS 480 360~720 480 <br />Interval time MINS /<40 20 <br />Initial evacuation KPa -60 -60±5.0 -60.0 <br />Humidity injection time MINS /0~4 3 <br />Humidity injection pressure KPa /-60±5.0 -59.8 <br />Humidity dwell MINS 30 30±2 30 <br />EO injection pressure KPa /-30±5.0 -29.7 <br />EO Weight (99.9%)Kg 15 15±0.5 15.0 <br />Humidity %RH /40~70 42.2~68.4 <br />Exposure temperature ℃50 45~55 47.1~52.7 <br />Exposure time MINS 480 480~500 480 <br />EO concentration mg/L 600 580~620 600 <br />Cleanings NUMBER 3 3 3 <br />Evacuation KPa -60 -60±5 -60.2 <br />Hold MINS 15 15~20 15 <br />Aeration time MINS 30 30~45 30 <br />Aeration temperature ℃//5~15 <br />Aeration time Days 7 7~8 7 <br /> We hereby certify that the above-mentioned goods have been sterilized by EO Gas according to the regulation of ISO 13485:2016 <br /> <br />WWW.TINADAVIES.COM 1.888.510.1388