Laserfiche WebLink
(c) Drain reservoir fill with clean distilled or demineralized water and run one <br /> 3 minute cycle (UNWRAPPED) at 1320C (2700F). <br /> (d) Drain reservoir and allow ster- <br /> ilizer to cool to room tempera- ---- r <br /> ture. Remove the trays, tray — J <br /> rack(1), and the tray plate(2). <br /> This is accomplished by �- <br /> grasping the tray rack (1) on I 2 <br /> both sides in the front and <br /> gently pulling outward. The I� <br /> tray rack and tray plate should 1 <br /> slide out of the chamber �i o <br /> Together. ,- <br /> (e) Wipe out the inside of the / 3 <br /> chamber being careful not to <br /> damage the heater element or <br /> the temperature and level 300 �► <br /> sensor components. Wipe off MA439000 <br /> .he trays, tray rack, and tray <br /> plate. <br /> (f) Re-install the tray rack (1) and tray plate (2) in the chamber as follows: <br /> Position the two rear posts of tray rack in rack holes (A) of tray plate. <br /> Then, hold front end of tray rack at approximately a 300 angle from the <br /> tray plate. Then, insert rear end of tray rack and tray plate as an assem- <br /> bly in chamber (3). Push tray rack and tray plate into chamber com- <br /> pletely. Re-install the trays. Operator <br /> (g) Refill the reservoir with clean distilled or demineralized water. Maintenance <br /> 2. Pe-form Pressure Relief Valve Check - The pressure relief valve must be <br /> checked each month by a qualified person to be sure that the relief valve is <br /> functioning properly: <br />