(c) Drain reservoir fill with clean distilled or demineralized water and run one <br />3 minute cycle (UNWRAPPED) at 132°C (270°F). <br />(d) Drain reservoir and allow ster- <br />ilizer to cool to room tempera- <br />ture. Remove the trays, tray <br />rack (1), and the tray plate (2). <br />This is accomplished by <br />grasping the tray rack (1) on <br />both sides in the front and <br />gently pulling outward. The <br />tray rack and tray plate should <br />slide out of the chamber <br />Together. <br />(e) Wipe out the inside of the <br />chamber being careful not to <br />damage the heater element or <br />the temperature and level <br />sensor components. Wipe off <br />.he trays, tray rack, and tray <br />olate. <br />MA439000 <br />(f) Re -install the tray rack (1) and tray plate (2) in the chamber as follows: <br />Position the two rear posts of tray rack in rack holes (A) of tray plate. <br />Then, hold front end of tray rack at approximately a 30° angle from the <br />tray plate. Then, insert rear end of tray rack and tray plate as an assem- <br />bly in chamber (3). Push tray rack and tray plate into chamber com- <br />pletely. Re -install the trays. <br />(g) Refill the reservoir with clean distilled or demineralized water. <br />2. Pe -form Pressure Relief Valve Check - The pressure relief valve must be <br />checked each month by a qualified person to be sure that the relief valve is <br />functioning properly: <br />