Laserfiche WebLink
1534 Willow Pass Road,Pittsburg,CA 94565-1701 <br /> 0 MCCAM PBELL ANALYTICAL INC. Telephone:925-252-9262 Fax:925-252-9269 <br /> W "When Quality Counts" <br /> O www.mccampbell.com:main@mccampbell.com <br /> leak of the sampling train during the Shut-In test, REPLACE the manifold and REPEAT the Shut-In <br /> test of the new manifold. <br /> 5.5.5 If no significant leak is detected; Valve I and purge can valve are reopened and soil permeability is <br /> noted. <br /> Note: Two criteria MUST be met;sample flow of at least 100mL/min AND borehole vacuum of-7.35 <br /> inHg are required to proceed. If these conditions are not met, damage to MAI equipment may occur <br /> at the expense of the sampler. Please see the "Sampling" section of this document for sampling <br /> procedure when the aforementioned criteria are not met. <br /> 5.6 Leak Test <br /> 5.6.1 The purpose of this test is to determine if ambient air is introduced into the sample during collection <br /> 5.6.2 The MAI Helium Shroud design allows for the Leak test to be performed while the Shut-In test is in <br /> progress. The Leak test is performed to verify a gas-tight connection between the borehole tubing <br /> andthe manifold as well as confirm a good grout"seal"prior to sample collection. <br /> 5.6.3 It is used for shallow wells to determine if there is high permeability between the sub-surface sample <br /> tubing and the ground surface. This test is useful only if sampling will be aborted if a borehole leak <br /> is detected. <br /> 5.6.4 While waiting the "one minute" time requirement of the Shut-In test, attach helium meter to end of <br /> blue tubing on Port 4. <br /> 5.6.5 Using 60mL syringe take a helium reading of the borehole; attach syringe to outlet of meter and give <br /> a quick pull on the plunger. (Pulling 60 mL of air through within a second is sufficient.) Repeat 2 <br /> times to get an accurate reading. <br /> 5.6.6 If shroud is charged to 20%(f2%)helium and meter reads< 1%helium in well,borehole seals pass <br /> DTSC<5%leak criteria. <br /> 5.7 Purge Volume Test <br /> 5.7.1 Note: There are several procedures for determining well volume and purging well volume. Please <br /> refer to the April 2012 DTSC Advisory section 4.2.3 — 4.3 for more information. An accepted <br /> method is listed below" however, this method may not be appropriate for all wells and may not <br /> always apply. <br /> 5.7.2 Turn 3-way Valve I so it is perpendicular to the manifold (pointing towards Port 4). Attach 60 mL <br /> syringe to end of blue tubing attached to outside of shroud box at Port 4. <br /> 5.7.3 Using the supplied 60mL syringe, pull suction on borehole sampling tube until plunger resistance is <br /> felt, noting final volume evacuated. The default purge volume is three times that of the well tubing <br /> volume(April 2012 DTSC Advisory <br /> 5.8 Sampling <br /> 5.8.1 After Shut-In test, Leak test, and purge volume procedures have been completed and helium <br /> concentration is maintained at 20% (f2%); ensure 3-way Valve I is parallel to manifold (pointing <br /> towards the gauges)and therefore allowing sampling from well. <br /> Page 7 of 9 <br />