Laserfiche WebLink
GUIXI CITY JINLONG TATTOO EQUIPMENT MANUFACTORY <br /> Xf File Number:JLTS-21-12 AMA,Ai Chamber No.: U21-12-07/CE <br /> ANIV.4#i;;j %1-rcM#�A-i11i&A t1•r'411HM Test Date:2021-12-07 <br /> Original Record of BL Growth Test after Sterilization <br /> 4�I" Customer:TATTOO EXPRESS SUPPLY INC <br /> Product Name:xxx cartridge needle <br /> /"i uJi�T Product Lot No. :21-12 <br /> A M H M Sterilization Date:21-12-07 AMA j Sterilization Lot.:JLWX2021-12-07-1 <br /> Wil"M R Test Quantity: 1000boxes t6t'A9Arl q Test Item: BI Growth Test Record <br /> no-i411 H M Test Date: 2021-12-07 Test Standard: ISO 11737-2:2005 <br /> Ar:Mlrfl'Ali�AP BI Growth Test Record Time Table <br /> t C i1 h-4: M-Vf M L�#TR-0Mh V1!A0: R1-1•I%1 1.1�IAA"4IN2 <br /> Strain:Bacillus Atrophaeus ATCC9372 BI Manufacturer:Hangzhou Future Biotech Co., Ltd <br /> BI 311 Lot No.: 191125 f f�&l Incubation Tempearture:36 `C <br /> to-rr.f>f I'A11A t<jW V)(Culture Time and BI Growth <br /> r Strip No.: 8 16 24 32 40 48 <br /> fc 1-10# -- -- -- — — -- <br /> 11# + + + + + + <br /> Result:(1-10#) ?1:i frfATcl'41rc, PHf'l::tffi'G c 11#> -6:1-rr'24-48,,1,11,1lr3�i M Itc. <br /> After 48 hours incubation, The(1-20#)are not bacteria growth, Positive(11#)bacteria growth well. <br /> !�r=i: Note: (-): �TrAft.represent negative,(+): cTMit represent positive. <br /> f-;Pq 'a it Detection Conclusion: <br /> ij#u7 g7E isf� p BI test of sterility: AM The results indicated PASSED <br /> J t i311 Test by: * L F Review by: *,H)j R M Rep :2 <br /> ,m any Stamp: <br />