Laserfiche WebLink
• Matrix Spike/Matrix Spike Duplicate (MS/MSD) Sample Performance <br /> • Field Duplicate Comparison <br /> • Detection Limits <br /> Summary of Findings: <br /> All samples were analyzed according to the accompanying chains of custody and all analytical holding <br /> times were met with one exception. Hexavalent chromium was analyzed outside of the 24 hour <br /> recommended holding time in 30 samples. Samples with hexavalent chromium detected were qualified <br /> as estimated (J); samples with no detectable hexavalent chromium concentrations were qualified as <br /> estimated non-detects (UJ) at the method detection limit. <br /> QC data were reviewed for laboratory and instrument precision and accuracy from LCS/LCSD recoveries <br /> and relative percent differences (RPDs) and MS/MSD sample recoveries and RPDs. All quality control <br /> elements were found to be within acceptable limits with the following exceptions: <br /> • The MS/MSD percent recoveries associated with the analysis of MW-2 were greater than the <br /> upper control limit for sulfate; the result was qualified as estimated (J) in the parent sample. <br /> • The MS/MSD percent recoveries and RPD associated with the analysis of MW-323 were not <br /> meaningful for sulfate. The concentration of sulfate in the parent sample was greater than four <br /> times the concentration used for the spike. Therefore, the criteria were not evaluated. <br /> • The MS/MSD percent recoveries and RPD associated with the analysis of TW-15 and MW-9 <br /> were not meaningful for manganese. The concentration of manganese in the parent samples <br /> was greater than four times the concentration used for the spike. Therefore, the criteria were not <br /> evaluated. <br /> Field sampling precision was also evaluated by using the calculated RPD between results reported for the <br /> field duplicate pairs, which are listed above. All RPD results were found to be within the acceptable limits <br /> for precision for all methods with the following exception: <br /> • The field duplicate pair MW-303 and MW-303 (Duplicate) had one RPD that was greater than the <br /> control limit for chromium; chromium was qualified as estimated (J) in the parent sample. <br /> No target analytes were detected in any method blank or field QC samples with the following exceptions: <br /> • Equipment Rinse and Field Blank, collected on April 24, 2012, were associated with the <br /> analysis of one sample and had TOC detected at 0.7 and 0.5 mg/L, respectively. EW-1 had TOC <br /> detected at a concentration that was greater than the reporting limit, but less than five times the <br /> blank concentration, and was qualified as estimated (J). <br /> • Equipment Rinse and Field Blank, collected on April 25, 2012, were associated with the <br /> analysis of four samples and had TOC detected at 1.0 and 0.8 mg/L, respectively, and TDS <br /> detected at 8.0 mg/L in both. TDS was detected in the associated samples at concentrations that <br /> were greater than five times the blank concentration and were not qualified. TOC was detected in <br /> four samples at concentrations that were greater than the reporting limit, but less than five times <br /> the blank concentration, and were qualified as estimated (J) in all four samples. <br /> • Equipment Rinse and Field Blank, collected on April 30, 2012, were associated with the <br /> analysis of four samples and had TDS detected at 6.0 and 8.0 mg/L, respectively, and TOC <br /> detected at 0.9 mg/L in both. TDS was detected in the associated samples at concentrations that <br /> were greater than five times the blank concentration and were not qualified. TOC was detected in <br /> Page 2of6 <br />