• September 14,2017 Lab ID : STK1751096 <br /> Live Oak Geo Environmental Customer : 3-16608 <br /> Quality Control - Inorganic <br /> Constituent Method Date/lD Type Units Cone. QC Data DQo Note <br /> Wet Chem <br /> urate+NiMte as 450ONO3F MS mg/L 28.04 35.4% 5-285 <br /> (STK1751081-002) MSD mg/L 28.04 34.7% 5-285 <br /> MSRPD m /L 28.04 2.1% 530.4 <br /> 450ONO3F 08131/17:213144JDD ICB mg(L 0.032 0.1 <br /> ICV mg/L 11.22 97.6% 90-110 <br /> CCB mg/L 0.040 0.1 <br /> CCV I m /L 1 11.22 1 97.2% 1 90-110 <br /> Definition <br /> ICV :Initial Calibration Verification-Analyzed to verify the instrument calibration is within criteria. <br /> ICB :Initial Calibration Blank-Analyzed to verify the instrument baseline is within criteria. <br /> CCV :Continuing Calibration Verification-Analyzed to verify the instrument calibration is within criteria. <br /> CCB :Continuing Calibration Blank-Analyzed to verify the instrument baseline is within criteria. <br /> MS Matrix Spikes-A random sample is spiked with a known amount of analyte.The recoveries are an indication of how that sample <br /> matrix affects analyte recovery. <br /> MSD :Matrix Spike Duplicate of MS/MSD pair-A random sample duplicate is spiked with a known amount of analyted.The recoveries <br /> are an indication of how that sample matrix affects analyte recovery. <br /> MSRPD :MS/MSD Relative Percent Difference(RPD)-The MS relative percent difference is an indication of precision for the preparation <br /> and analysis. <br /> DQO :Data Quality Objective-This is the criteria against which the quality control data is compared. <br /> • <br /> • <br /> Page 7 of 7 <br />