C <br /> SIERRA <br /> SES STANDARD OPERATING PROCEDURE <br /> OVM READINGS <br /> SES uses an organic vapor meter (OVM) to determine the presence or absence of volatile <br /> organic compounds(VOCs),including benzene,toluene,ethylbenzene, and xylenes in soil samples <br /> chosen for field screening. The OVM uses a photoionization detector (PID) and is calibrated daily <br /> to 100 parts per million of 1-liter of isobutylene. The OVM, which measures in parts per million <br /> by volume (ppmv), is used for qualitative, not quantitative, assessment because the correlation <br /> between the volume measurements of the OVM and the weight measurements of the laboratory <br /> instruments is not well defined. <br /> A field screen sample is obtained from the brass tube immediately above or below the brass <br /> tube containing the sample selected for possible analysis. The soil to be screened is removed from <br /> the brass tube, and is placed in a pre-cleaned brass tube with aluminum foil and a polyethylene <br /> cap on one end. The brass tube is loosely filled to approximately 1/2 full. Another square of <br /> aluminum foil is placed on the open end and a polyethylene cap with crossed slits is placed over <br /> it. <br /> The field screen sample is allowed to temperature equilibrate for approximately 15 to 30 <br /> minutes in the sun, allowing any VOCs which might be present in the soil to volatilize out into the <br /> brass tube's headspace. The OVM nozzle is then placed inside the sealed brass tube, through the <br /> slits in the cap, in order to measure the VOCs present, if any, in the headspace. The nozzle <br /> should remain inside the brass tube for approximately 15 to 30 seconds or until the maximum <br /> reading has been recorded on the OVM readout panel. <br /> The depth from which the sample came and the corresponding OVM reading is recorded <br /> on the original field log sheet. Field observations, OVM and (odor and staining) readings are used <br /> in determining which soil samples are to be analyzed in the laboratory. <br /> OVM.SOP <br />