8.2 Setting Up For A Run (Ready Mode) <br /> KEY DISPLAY NOTES <br /> 1. NEXT LOGGING DISABLED This screen indicates the <br /> logging status.To enable <br /> the logging function,key <br /> ENTER. Toggling of the <br /> on/off logging status can <br /> be done by keying ENTER. <br /> 2. ENTER LOG INTRVL 600s This message indicates that logging is <br /> TAG#: 4 enabled. Example is for 10-min log <br /> period,selected through the PC (see <br /> Section 9.0), and next free tag is#4. <br /> 3. NEXT ALARM: OFF This screen indicates the <br /> alarm status. Keying ENTER <br /> repeatedly toggles through <br /> the 3 alarm modes: <br /> 4. ENTER ALARM: INSTANT This enables the alarm based <br /> LEVEL:1.50 mg/m3 on the real-time concentration. <br /> The level (e.g. 1.50 mg/m3) <br /> must be set on the PC. <br /> 5. ENTER ALARM: STEL This enables the alarm based <br /> LEVEL:0.50 mg/m3 on the 15-min STEL value. <br /> The level (e.g. 0.50 mg/m3) <br /> must be set on the PC. <br /> 6. NEXT ANALOG OUTPUT: This screen indicates the analog signal <br /> DISABLED output status. Keying ENTER will <br /> enable the analog output. Toggling <br /> the analog output on/off can be done <br /> by keying ENTER: <br /> 7. ENTER ANALOG OUTPUT: This enables the analog output. The <br /> 0 - 0.400 mg/m3 concentration range (e.g., 0 - 0.400 <br /> Mg/M3) must be set on the PC. <br /> 8. NEXT CAL FACTOR: 1.00 This screen displays the <br /> DIS AVG TIME 10s calibration factor and the <br /> display averaging time. <br /> Both values can be edited via <br /> PC. <br /> 19 <br />