Laserfiche WebLink
9. NEXT BATTERY LEFT 83% This screen displays the <br /> MEMORY LEFT 96% remaining battery charge,and the <br /> remaining percentage of free memory. <br /> 10. NEXT CONNECT TO PC When this screen has been <br /> selected,the operating <br /> parameters can be edited <br /> and/or the logged data can <br /> be downloaded via the PC <br /> (see Section 9.0). If NEXT <br /> is keyed again,the screen <br /> returns to RUN/READY: <br /> 11. NEXT START RUN:ENTER The instrument is now <br /> READY:NEXT ready to run following the <br /> procedure in section 8.3. <br /> 8.3 Measurement Run Procedure <br /> KEY DISPLAY NOTES <br /> 1. ENTER LOGGING DISABLED or,if logging was enabled: <br /> LOG INTRVL 600s Logging status will be displUed <br /> TAG#:4 for 3 seconds. <br /> CONC*0.047 mg/m3 After a 3-second delay,the concent- <br /> TWA 0.039 mg/m3 ration screen appears (values <br /> shown here are examples). <br /> CONC is the real-time and TWA <br /> is the time-averaged concent- <br /> ration. <br /> oncentration.The*ap ep ars only <br /> logging has been enabled. <br /> 2. EXIT TERMINATE RUN? To terminate the current run <br /> Y:ENTER N:EXIT and return to the Ready Mode, <br /> key ENTER. To continue the <br /> run,key EXIT. <br /> 3. EXIT CONC*0.047 mg/m3 Keying NEXT successively <br /> TWA 0.039 mg/m3 scrolls the display to show <br /> various run values (elapsed <br /> run time,maximum,STEL, <br /> etc.). Keying EXIT from any <br /> of those screens returns to <br /> the concentration display. <br /> 20 <br />