Laserfiche WebLink
DEMENNO / KERDOON <br /> ICP P , E 4300DV <br /> OA REPORT FORM <br /> DATE : 5/29/2018 ANALYST: JL <br /> ANALYTE ICV % REC , ICB ICSA % REC . ICSAB % REC . LCS % REC . CCV % REC CCB <br /> Pb 1014 101 < 25 512 102 517 103 107 107 981 98 <25 <br /> Cu 1006 101 <25 509 102 510 102 104 104 961 96 <25 <br /> Zn 1039 104 <25 525 105 526 105 114 114 993 99 <'25 <br /> TI 1008 101 <25 505 101 507 101 102 102 979 98 < 25 <br /> As 1023 102 <25 522 104 522 104 111 111 983 98 <25 <br /> Cr 1024 102 <25 517 103 517 103 106 106 981 98 <25 <br /> Ni 1017 102 < 25 514 103 515 103 105 105 974 97 < 25 <br /> Cd 1027 103 < 25 520 104 520 104 105 105 987 99 X25 <br /> Ag 1010 101 <25 507 1041 514 103 105 105 985 99 <25 <br /> Ba 1021 102 <25 516 103 519 104 104 104 987 99 <25 <br /> Se 1020 102 < 25 510 102 503 101 100 100 974 97 <25 <br /> Be 1024 102 < 25 520 104 520 104 104 104 996 100 <25 <br /> Mo 1016 102 <25 520 104 522 104 107 107 983 98 <25 <br /> Sb 996 9906 < 25 511 102 514 103 102 102 966 97 < 25 <br /> V 1015 102 <25 509 102 509 102 106 106 967 97 <25 <br /> Co 1025 103 <25 515 103 516 103 106 106 980 98 <25 <br /> Sn 1020 102 < 25 533 107 523 105 108 108 983 98 <25 <br /> ANALYTE CCV % REG CCB CCV % RECO CCB CCV % REC . CCB CCV % REC . CCB <br /> Pb 999 100 <25 <br /> Cu 983 98 <25 <br /> Zn 1015 102 < 25 <br /> TI 989 99 <25 <br /> As 1000 100 <25 <br /> Cr 1005 101 <25 <br /> Ni 995 100 <25 <br /> Cd 1006 101 <25 <br /> Ag 1010 101 < 25 <br /> Ba 1011 101 <25 <br /> Se 990 99 <25 <br /> Be 1023 102 < 25 <br /> Mo 1000 100 < 25 <br /> Sb 980 98 < 25 <br /> V 993 99 <25 <br /> Co 1003 100 <25 <br /> Sn 1000 100 < 25 <br /> ICV, - INITIAL CALIBRATION VERIFICATION ICSA . - INTERFERENT CHECK STANDARD SOLUTION 0. 5 PPM <br /> CCV, - CONTINUING CALIBRATION VERIFICATION ICSAB. - INTERFERENT CHECK STANDARD SOLUTION 0. 5 PPM <br /> ICB , - INITIAL CALIBRATION BLANK % R CRITERIA ON ICSH 80420 <br /> CCB , - CONTINUING CALIBRATION BLANK % R CRITERIA ON ICSAB 80420 <br /> % CRITERIA ON ICV 90- 110 LCS• LABORATORY CONTROL SAMPLE( 100 PPB) <br /> % CRITERIA ON CCV 80420 % R CRITERIA ON LCS 85- 115 <br /> DATE : <br /> CHECKED BY : <br /> REVIEWED BY : DATE: � <br /> i <br />