Laserfiche WebLink
r] <br />Lodi Memorial Hospital <br />975 S. Fairmont Avenue <br />Lodi, CA 95240 <br />LT1701 V "Ile] V Q =1 <br />Control # <br />1446 <br />Description <br />STERILIZER, WASTE <br />Manufacturer <br />Mark Costello <br />Model <br />AS -46 <br />Serial <br />LAAT-975 <br />Cost Center <br />8460 Plant Maintenance <br />Equipment <br />Notes <br />Instructions <br />0 <br />Maintenance <br />Building <br />Lodi Memorial Hospital <br />Location <br />OUTSIDE <br />Class <br />WST STIER <br />Issue Date <br />5/1/2006 Due 6/30/2006 <br />Assigned to <br />Risk <br />2 Critical <br />Safety <br />Inspection <br />Performance Allow the sterilizer to stablize at the set temperature. Using a calibrated thermometer, verify the <br />Inspection temperature displayed on the chart recorder is within +0/ -5 degrees F. Record temperature accuracy <br />in the service notes. <br />Forward copy of of report to Donna. <br />Preventive CLEAN AND CHECK OVERALL CONDITION <br />Maintenance GREASE DOOR AND LATCH <br />GREASE CART CASTERS <br />REMOVE DEBRI FROM DRAIN INSIDE VESSEL (DO NOT ENTER!) <br />REMOVE STRAINER AND CLEAN <br />VERIFY STEAM PRESSURE AND RECORD 38 LBS. MINIMUM <br />Parts <br />Replacement <br />Service Date Tech Hr Service Notes <br />616/2006 Duncan 1.50 Unit was checked with a calibrated thermometer, and <br />temperature was within + - 2 degrees of actual recorded <br />temperature. <br />Complete W Closed By Aqngt Date 6/6/2006 Hours <br />