Initiai Start Up <br /> Make sure the Batte[y and the aLaba are properly <br /> connected. <br /> 15 minute warm up. <br /> t.It A- <br /> Activate the "Auto Store' option. (section 0.a.) <br /> Disable the "Dispiay Thresholds." (section 1.3.b.) <br /> Note: Screen should read, "enable display thresholds." <br /> Operations Checks <br /> Perform Eneroy Calibration only when instrument has <br /> moved from one site to another, every month or system <br /> fails resolution check (use lead cover shield). (section i+.1.) <br /> Change source times to 60 counts. (section 11.2.) <br /> Perform Energy Calibration Check (use lead cover) <br /> Fill out Daily instrument ChecKout Worksheet. (section i1.3.) <br /> Perform Resoiution Check (use pure FE standard). <br /> (section 11.4.) <br /> Intensity Criteria <br /> Mn <0.003 <br /> Fe >0.9C0 <br /> Co <0.003 <br /> 'Perform Energy Calibration if Intensity Criteria can not bE <br /> attained. <br /> Perform Backaround Check (use teflon blank; section <br /> 'Results should be less than 3 times the standard <br /> 'Perform Acquire Background rf Background Check can <br /> not be attained. <br />