C. Maintenance <br />9 <br />Note: Strict programming of the control electronics oj all San-I-Pak units prevents the discharge of <br />unslerilized waste. San-l-Pak employs parametric motoring, which is superior to Biological Tests. <br />The computer has a built-in steam table that enables constant comparison of vessel temperature with <br />vessel pressure. Other fail-safes such as time to vacuum, vacuum level, time to temperature, droop <br />control and vent time exceed conventional testing. In the unusual occurrence of any discrepancy in <br />the programmed parameters, the process will be halted and then after correction cycle must be <br />restarted. <br />Maintenance of the autoclaves is performed consistent with the autoclave manufacturer’s <br />recommendations, which include the following schedule of maintenance activities and planned <br />frequency. <br />1710 Incubator -Getinge model <br />Dox Matrix Printer-Model-CBM-9IO-Type 11 <br />Serial Number-SW20127000 <br />The biological ampoule contained within the AccuFast Biological Test Pack, which Includes a paper <br />strip containing G. stearothermophilus inside an ampoule of specially formulated growth medium, <br />should be incubated at approximately 60°C. for 10 hours. After incubation, the ampoule visually <br />indicates growth or no growth of the indicator spore. No color change (any growth) equals complete <br />sterilization. Sterilization failure is noted by positive growth of the bacteria which turns the medium <br />color from purple (PASS) to yellow (FAIL). <br />color change is not a true confirmation of fail or pass as the incubator visual change is the true <br />indication of pass and fail for each load. <br />Controls H