Laserfiche WebLink
t �T' <br /> venue South €fD7,Pacheco CA 94553 5560 <br /> �4�McC�ampbe�IlAnalyt>cal Inc one 925-798-1620 Fax 92s 798 1622ccampbell cam Email mam(r�mccampbelt coin <br /> QC SUMMARY REPORT FOR SW802IB/8015Cm <br /> Matrix W WorkOrder 0212472 <br /> EPA Method SW8021 B18015CM Extraction SW6030B BatchlD 5501 Spiked Sample ID -0212472-002A <br /> Sample Spiked MS* MSD* MS-MSD* LCS LCSD LCS-LCSD Acceptance Criteria(%) <br /> Compound , <br /> ' pg/L VgIL %Rec %Rec %RPD %Rec %Rec % RPD Low High <br /> TPH(gas) ND 60 105 105 0 620 101 993 203 80 120 <br /> ' MTBE ND 10 103 102 1 60 943 934 0 898 80 120 <br /> Benzene ND 10 100 998 0 140 97 7 974 0 291 80 120 <br /> Toluene ND 10 926 91 1 78 101 99 7 151 80 120 <br /> ' Ethylbenzene ND 10 962 956 0 642 95 6 959 0320 80 120 <br /> Xylenes ND 30 927 89 404 993 993 0 80 120 <br /> ' %SS 107 100 995 994 0 134 945 945 0000741 80 120 <br /> All target compounds in the Method Blank ofthis extraction batch were ND less than the method RI,with the following exceptions <br /> ' NOME <br /> 1 � i <br /> 1 <br /> MS=Matrix Spike MSD=Matrix Spike Duplicate LCS=Laboratory Control Sample LCSD=Laboratory Control Sample Duplicate RPD=Relative Percent <br /> ' Deviation <br /> NIA=not enough sample to perform matrix spike and matrix spike duplicate <br /> NR=analyte concentration in sample exceeds spike amount for sod matrix or exceeds 2x spike amount for water matrix or sample diluted due to high matrix or <br /> ' anatyte content <br /> %Recovery=100'(MS-Sample)I(Amount Spiced) RPD=100*(MS--MSD)I(MS+MSD)'2 <br /> -MS and 1 or MSD spike recoveries may not be near 100%or the RPDs near 0%if a)the sample is inhomogenous ANP contains significant concentrations of <br /> ' anatyte relative to the amount spiked or b)if that specific sample matrix interferes with spike recovery <br />