Laserfiche WebLink
LABORATORY CONTROL SAMPLE EVALUATION REPORT <br />GC Volatiles <br />Client Lot #...: <br />ElKlS0233 <br />Work Order #...: <br />EN5QA1AC <br />Matrix.........: SOLID <br />LCS Lot -Sample#: <br />E1K160000-406 <br />- <br />Prep Date......: <br />11/15/01 <br />Analysis <br />Date..: <br />11/15/01 <br />Prep Batch #...: <br />1320406 <br />Analysis <br />Time..: <br />15:02 <br />Dilution Factor: <br />1 <br />Instrument ID..: <br />G15 <br />Analyst ID.....: <br />018568 <br />PERCENT <br />RECOVERY <br />PARAMETER <br />RECOVERY <br />LIMITS <br />METHOD <br />Benzene <br />108 <br />(80 <br />- 125) <br />SW846 <br />8021B <br />Toluene <br />112 <br />(80 <br />- 125) <br />SW846 <br />8021B <br />Ethylbenzene <br />98 <br />(80 <br />- 120) <br />SW846 <br />8021B <br />Xylenes (total) <br />110 <br />(80 <br />- 120) <br />SW846 <br />8021B <br />m -Xylene & p -Xylene <br />110 <br />(80 <br />- 120) <br />SW846 <br />8021B <br />o -Xylene <br />109 <br />(80 <br />- 120) <br />SW846 <br />8021B <br />PERCENT <br />RECOVERY <br />SURROGATE <br />RECOVERY <br />LIMITS <br />a,a,a-Trifluorotoluene <br />106 <br />(60 - <br />130) <br />(TFT) <br />NOTE (S) <br />Calculations are performed before rounding to avoid round -off errors in calculated results. <br />Bold print denotes control parameters <br />• <br />