n <br />►' r <br />MANUAL TITLE: <br />ENVIRONMENTAL SERVICES <br />POLICY NUMBER <br />EVS 181.1 <br />SECTION IV SPECIFIC POLICIES AND <br />PROCEDURES TO THE DEPARTMENT <br />ISSUE 5/04 <br />SUBJECT: SPORE TEST POLICY <br />REVIEWED 5/04 <br />APPROVED BY MONICA GRAY <br />PAGE 1 OF 3 <br />REFERENCE <br />POILCY: Evidence of an effective sterilization process will be documented. <br />Rational: Load is the process by which a load is monitored and released based on <br />the results of the biological indicator. All personnel performing sterilization activities will <br />follow the procedure for incubation and reading of all biological indicators. <br />1. Always leave the 3MTM AttestTM Auto -reader plugged in. If found unplugged, allow <br />30 minutes warm up time. <br />2. Utilize biological indicator 1296 Rapid Readout tubes. <br />3. Always wear Personal Protective Equipment (PPE), when operating the machine. <br />• Safety Goggles <br />• Heat Resistant Gloves <br />4. Load the Sani-Pak chamber half way with red bags enclosed in autoclavable bags. <br />(Each autoclave bag must be punctured for ventilation). Place biological indicator test <br />pack on top bag in the center of the load. Run cycle. <br />5. The sterilized tube test pack is removed from the SaniPak and the test pack is allowed <br />to cool prior to handling and processing. <br />6. Be sure to use tubes from the same manufacturer, date and lot number for each test. <br />(Example: lot # 2006-01 AK, which means the lot expires January 2006). <br />