7.3.2.2 A newly prepared chamber working standard is rejected <br />unless the relative response factors for all eight <br />halogenated hydrocarbons of interest fall within <br />+/- 10% of the historically established mean values. <br />7.5.2 The mean -square error due to lack of fit about the <br />regression line is compared to the total mean -square <br />error of the independent replicates about their <br />individual means. The calibration is accepted if <br />the F -ratio is less than the 95% rejection limit. <br />7 <br />7.3.3 <br />A working chamber standard is checked at least every <br />three months for conformity to criteria 7.3.2.1 and <br />7.3.2.2. <br />7.3.3.1 <br />A new standard is prepared as frequently as required as <br />determined by the above mentioned criteria. <br />7.3.3.2 <br />Any reports generated after the standard ceases to be <br />demonstratively within the established tolerances shall <br />contain a cautionary explanation. <br />7.4 <br />The gas chromatograph is calibrated periodically. <br />7.4.1 <br />Calibration factors are determined on the basis of the <br />mean values of the previous calibration runs which meet <br />the criteria of 7.4.3. <br />7.4.2 <br />Each day a calibration check is performed using the <br />Pfaudler chamber standard to span the instrument. <br />7.4.3 <br />If the response for each compound of interest is within <br />10% of the established calibration value, the established <br />calibration factors are retained. <br />7.4.4 <br />The calibration check is repeated if the response of the <br />instrument has changed by more than 10% from the established <br />values. <br />7.4.5 <br />If the response is still out of tolerance, a quality <br />assurance report is submitted, remedial action is initiated, <br />and new calibration factors calculated. <br />7.4.6 <br />Blank samples shall be analyzed daily after the calibration <br />is completed and, whenever necessary, between samples. <br />7.5 <br />The linearity of the instrument is checked periodically. <br />7.5.1 <br />A gas chromatographic multipoint linearity check is <br />performed annually with standards of at least four <br />different concentrations and four replicate runs for <br />each concentration. The concentrations should include <br />the anticipated range of sample concentrations above <br />the limit of d6tection. <br />7.5.2 The mean -square error due to lack of fit about the <br />regression line is compared to the total mean -square <br />error of the independent replicates about their <br />individual means. The calibration is accepted if <br />the F -ratio is less than the 95% rejection limit. <br />7 <br />