7.5.3 <br />A repeated multipoint calibration should not differ <br />from the previous calibration by more than 10%. <br />7.5.4 <br />Any region of concentration that deviates more than <br />5% from the least -square line is considered nonlinear. <br />7.5.5 <br />Data is reported only for compounds whose concentra- <br />tions lie in the linear range. <br />7.6 <br />Limits of detection are established. <br />7.6.1 <br />The limit of detection (LOD) is based on three standard <br />deviations (SD) of runs near the LOD (within 10 SD of <br />the LOD, Winefordner and Long, 1983). <br />7.6.2 <br />The LOD should be determined at least on an annual <br />basis. <br />7.6.3 <br />If the instrument response changes by more than 15%, the <br />instrument must be checked and the LOD redetermined. <br />7.6.4 <br />The presence in a sample of a very large adjacent <br />peak will often raise the LOD in the sample. <br />7.7 <br />Analytical instruments have quality control procedures. <br />7.7.1 <br />Column conditions are checked periodically and as needed. <br />7.7.1.1 <br />All GC accessible parameters is logged when a <br />column is first installed. These parameters are <br />checked daily and recorded on integrator reports. <br />7.7.1.2 <br />The efficiency and resolution of the column are <br />checked every month. If the tests show more than a <br />10% change, the column is replaced. <br />7.7.1.3 <br />If the headpressure required to maintain a specified <br />flow through the column increases by more than 100%, <br />the column is replaced. <br />7.7.1.4 <br />If the drift of retention times of peaks results <br />in peak misidentification, all instrument parameters <br />are checked. <br />7.7.2 <br />Replicate analyses are a quality control procedure. <br />7.7.2.1 <br />A duplicate analysis is performed on at least one <br />sample per day. <br />7.7.2.2 <br />If the duplicate analysis (replicate) differs by more <br />than 20%, and if the concentration of the sample is <br />higher than 3X LOD, then an additional analysis is <br />performed. <br />8 <br />