Laserfiche WebLink
Troubleshooting Guide - (Continued) <br /> Problem Possible Cause Solution <br /> Sterilization failure sterilization conditions reload the sterilizer in <br /> evidence from process were not present at the accordance with Guidelines <br /> monitor(chemical location of the indicator. for Loading Trays. If the <br /> indicator, biological problem recurs, take the <br /> indicator, etc.) unit out of service and <br /> contact an authorized <br /> service representative (see <br /> Calling For Service later in <br /> this manual). <br /> insufficient air removal, low take the unit out of service <br /> temperature, or low and contact an authorized <br /> pressure. service representative (see <br /> Calling For Service later in <br /> this manual). <br /> indicator is out of date, is use an indicator, <br /> inappropriate for sterilizer appropriate for the load and <br /> cycle, or has cycle selected, from a fresh <br /> malfunctioned. supply that has been stored <br /> properly. Contact the <br /> indicator manufacturer for <br /> additional information on <br /> proper selection, use, <br /> storage, and potential <br /> misapplication or <br /> malfunction. <br /> Door gasket leaks. door gasket is damaged or allow sterilizer to cool, then <br /> dirty. clean or replace door <br /> gasket(see Remove and <br /> Clean Door Gasket under ' ' <br /> Quarterly maintenance in Maintenance <br /> this manual). <br /> sterilizer is not level. level the sterilizer. <br /> Packs not dry sterilizer is overloaded. reload the sterilizer in <br /> accordance with Guidelines <br /> for Loading Trays. If the <br /> problem recurs, take the <br /> unit out of service and <br /> contact an authorized <br /> service representative (see <br /> Calling For Service later in <br /> this manual). <br /> chamber filter is clogged. allow sterilizer to cool; then <br /> clean or replace filter(see <br /> Cleaning Chamber Filter <br /> under Monthly maintenance <br /> in this manual). <br />