Error Code Guide - (Continued) <br />Problem Possible Cause <br />Error Code 5 sterilizer overloaded with <br />(excessive pressure) gauze or linen packs. <br />Unit is out of calibration or <br />bellows is not functioning <br />properly. <br />Solution <br />reload the sterilizer in <br />accordance with Guidelines <br />for Loading Trays. If the <br />problem recurs, take the <br />unit out of service and <br />contact an authorized <br />service representative (see <br />Calling For Service later in <br />this manual). <br />allow sterilizer to cool <br />completely. Restart <br />sterilizer. If unit has errors <br />again, unplug unit and <br />contact an authorized <br />service representative (see <br />Calling For Service later in <br />this manual). <br />Error Code 6 unit is out of calibration. allow sterilizer to cool <br />(excessive temperature) completely. Restart <br />sterilizer. If unit has errors <br />again, unplug unit and <br />contact an authorized <br />service representative (see <br />Calling For Service later in <br />this manual). <br />Error Code 7 <br />(pressure low during <br />cycle) <br />leaking door gasket <br />internal leak in unit <br />plumbing. <br />allow sterilizer to cool; then <br />clean or replace door <br />gasket (see Remove and <br />Clean Door Gasket under <br />Quarterly maintenance in <br />this manual). <br />allow sterilizer to cool <br />completely. Restart <br />sterilizer. If unit has errors <br />again, unplug unit and <br />contact an authorized <br />service representative (see <br />Calling For Service later in <br />this manual). <br />Error Code 8 door did not open at end of make sure the tray rack is <br />(door malfunction) cycle. pushed all the way back into <br />the chamber and is not <br />making contact with door. <br />