Laserfiche WebLink
� SAN1 fPA'K 'OPERATION ,ONTINUED . . . <br /> STEP 712 Slide open the compartment door for general waste , and remove <br /> the "Thermo- Indicator" strip : The lettering "OK" must be <br /> black in color to indicate a satisfactory sterilization . ShaId <br /> the latter " K" not be black in color , cease the procedure at <br /> this point and notify your supervisor . Should the letters be <br /> black in color , continue the operation with the next step . <br /> STEP 713 Slide the compartment door to general waste closed and lock . <br /> STEP 714 Depress the blue " Discharge" button once more and the sterilizer <br /> chamber will return to the original location . <br /> STEP 715 Depress the single " Compact" button , adjacent to the general <br /> waste door , and the sterilized waste will be moved into the <br /> receiver and compacted . <br /> STEP 716 Once the contaminated waste has been properly disposed of, and <br /> the unit is to be shut down , open both chambers individually <br /> and , with the " cleaning wand" , wash the interior of both the <br /> "Biohazare" and "General Waste" units . Now , wash down the <br /> exterior of the Sani -Pak system . <br /> STEP 717 Record the following information into the appropriate log <br /> book in the Environmental Services office area : <br /> - Heat/Temperature : Taken from the recorder ( 7-day clock ) , <br /> with each load . <br /> - Pressure ( P . S . I . ) : Taken from pressure guage during cycl - <br /> with each load . <br /> - Test Strips : Taken from waste chamber upon completion <br /> of cycle , with each load . <br /> - Bacteria Test : Run once per week with a load to be <br /> sterilized . Results recorded upon receipt <br /> from Bacteriology . <br /> NOTE: Thermometer calibration of the unit is to be done once per.year. <br /> err ari , M.0. <br /> Chairman <br /> Infection Control Committee <br /> JL : js <br /> 8/ 1985 <br />