Laserfiche WebLink
Loads Including Instruments <br /> The maximum M9 sterilization load shall be 140 instruments. If a load surpass- <br /> es these limits, we recommend dividing the load and running multiple cycles. <br /> Listed below are the maximum recommended loads for each tray: <br /> Maximum Capacities <br /> Load Type M9 Largeey Deep M9 Small Tray Sterilizer Total <br /> Tr <br /> Solid Items 42 instruments-1100 28 instruments-700 140 instruments-3.6 <br /> grams(2.4 lbs.) or grams(1.6 lbs.) or kg (8.0 lbs) or <br /> Liquids M 90-8 ml (0.27 oz) 90-8 ml (0.27 oz) 90-8 ml (0.27 oz) <br /> containers or containers or containers or <br /> Packs W 1080 cu.cm up.to 2.5 750 cu.cm up to 2.5 3660 cu.cm up to 2.5 <br /> cm thick(66 cu.in.up cm thick(48 cu.in.up cm thick(228 cu. in. <br /> to 1 in. thick) to 1 in.thick) up to 1 in.thick) <br /> t Not recommended for sterilization of liquids intended for direct patient con- <br /> tact <br /> $ Packs to have a minimum of 114 in. (6.3 mm)space between each other and <br /> away from all sterilizer surfaces. <br /> Operation <br />