Loads Including Instruments <br />The maximum M9 sterilization load shall be 140 instruments. If a load surpass- <br />es these limits, we recommend dividing the load and running multiple cycles. <br />Listed below are the maximum recommended loads for each tray: <br />Maximum Capacities <br />Load Type <br />M9 Largeey Deep <br />Tra <br />M9 Small Tray <br />Sterilizer Total <br />Solid Items 42 instruments -1100 <br />28 instruments - 700 <br />140 instruments - 3.6 <br />grams (2.4 lbs.) or <br />grams (1.6 lbs.) or <br />kg (8.0 lbs) or <br />Liquids M 90 - 8 ml (0.27 oz) <br />90 - 8 ml (0.27 oz) <br />90 - 8 ml (0.27 oz) <br />containers or <br />containers or <br />containers or <br />Packs W 1080 cu. cm up.to 2.5 <br />750 cu. cm up to 2.5 <br />3660 cu. cm up to 2.5 <br />cm thick (66 cu. in. up <br />cm thick (48 cu. in. up <br />cm thick (228 cu. in. <br />to 1 in. thick) <br />to 1 in. thick) <br />up to 1 in. thick) <br />t Not recommended for sterilization <br />of liquids intended for direct patient con- <br />tact <br />$ Packs to have a minimum of 1/4 in. <br />(6.3 mm) space between each other and <br />away from all sterilizer surfaces. <br />